Determination of small field output factors and correction factors using a Monte Carlo method for a 1000 MU/min CyberKnife® system equipped with fixed collimators
Tóm tắt
Từ khóa
Tài liệu tham khảo
Alfonso, 2008, A new formalism for reference dosimetry of small and nonstandard fields, Med. Phys., 35, 5179, 10.1118/1.3005481
Aspradakis, 2010
Bassinet, 2010, Characterization of 7LiF: Mg,Ti TLD micro-cubes, Radiat. Meas., 45, 646, 10.1016/j.radmeas.2009.12.005
Bassinet, 2013, Small fields output factors measurements and correction factors determination for several detectors for a CyberKnife® and linear accelerators equipped with microMLC and circular cones, Med. Phys., 40, 071725, 10.1118/1.4811139
Cranmer-Sargison, 2011, Implementing a newly proposed Monte Carlo based small field dosimetry formalism for a comprehensive set of diode detectors, Med. Phys., 38, 6592, 10.1118/1.3658572
Czarnecki, 2013, Monte Carlo calculated correction factors for diodes and ion chambers in small photon fields, Phys. Med. Biol., 58, 2431, 10.1088/0031-9155/58/8/2431
Das, 2008, Small fields: non equilibrium radiation dosimetry, Med. Phys., 35, 206, 10.1118/1.2815356
Francescon, 2008, Total scatter factors of small beams: a multidetector and Monte Carlo study, Med. Phys., 35, 504, 10.1118/1.2828195
Francescon, 2011, Calculation of kQclin,Qmsrfclin,fmsr for several small detectors and for two linear accelerators using Monte Carlo simulations, Med. Phys., 38, 6513, 10.1118/1.3660770
Francescon, 2012, Monte Carlo simulated correction factors for machine specific reference field dose calibration and output factor measurement using fixed and iris collimators on the CyberKnife system, Phys. Med. Biol., 57, 3741, 10.1088/0031-9155/57/12/3741
Heydarian, 1996, A comparison of dosimetry techniques in stereotactic radiosurgery, Phys. Med. Biol., 41, 93, 10.1088/0031-9155/41/1/008
Huet, 2012, Characterization and optimization of EBT2 radiochromic films dosimetry system for precise measurements of output factors in small fields used in radiotherapy, Radiat. Meas., 47, 40, 10.1016/j.radmeas.2011.10.020
Lechner, 2013, Detector comparison for small field output factor measurements in flattening filter free photon beams, Radiother. Oncol., 109, 356, 10.1016/j.radonc.2013.10.022
Moignier, 2014, Determination of the kQclin,Qmsrfclin,fmsr correction factors for detectors used with an 800 MU/min CyberKnife® system equipped with fixed collimators and a study of detector response to small photon beams using a Monte Carlo method, Med. Phys., 41, 10.1118/1.4881098
Morin, 2013, A comparative study of small field total scatter factors and dose profiles using plastic scintillation detectors and other stereotactic dosimeters: the case of the CyberKnife, Med. Phys., 40, 10.1118/1.4772190
Pantelis, 2012, On the output factor measurements of the CyberKnife iris collimator small fields: experimental determination of the kQclin,Qmsrfclin,fmsr correction factors for microchamber and diode detectors, Med. Phys., 39, 4875, 10.1118/1.4736810
Paskalev, 2003, Physical aspects of dynamic stereotactic radiosurgery with very small photon beams (1.5 and 3 mm in diameter), Med. Phys., 30, 111, 10.1118/1.1536290
Ralston, 2012, Small field diode correction factors derived using an air core fibre optic scintillation dosimeter and EBT2 film, Phys. Med. Biol., 57, 2587, 10.1088/0031-9155/57/9/2587
Scott, 2008, Using a Monte Carlo model to predict dosimetric properties of small radiotherapy photon fields, Med. Phys., 35, 4671, 10.1118/1.2975223
Scott, 2012, Characterizing the influence of detector density on dosimeter response in non-equilibrium small photon fields, Phys. Med. Biol., 57, 4461, 10.1088/0031-9155/57/14/4461