Small fields output factors measurements and correction factors determination for several detectors for a CyberKnife® and linear accelerators equipped with microMLC and circular cones

Medical Physics - Tập 40 Số 7 - 2013
C. Bassinet1, C. Huet1, S. Derreumaux1, G. Brunet2, M. Chéa3, M. Baumann1, T. Lacornerie4, S. Gaudaire‐Josset2, F. Trompier1, P. Roch1, G. Boisserie3, I. Clairand1
1Institut de radioprotection et de sûreté nucléaire (IRSN), BP17, 92262 Fontenay-aux Roses Cedex, France
2Institut de Cancérologie de l'Ouest René Gauducheau, bd Jacques Monod, 44805 Saint Herblain Cedex, France
3Groupe Hospitalier Pitié-Salpêtrière, 47/83 bd de l'Hôpital, 75651 Paris Cedex 13, France
4Centre Oscar Lambret, 3, rue Frédéric Combemale, BP 307, 59020 Lille Cedex, France

Tóm tắt

Purpose:The use of small photon fields is now an established practice in stereotactic radiosurgery and radiotherapy. However, due to a lack of lateral electron equilibrium and high dose gradients, it is difficult to accurately measure the dosimetric quantities required for the commissioning of such systems. Moreover, there is still no metrological dosimetric reference for this kind of beam today. In this context, the first objective of this work was to determine and to compare small fields output factors (OF) measured with different types of active detectors and passive dosimeters for three types of facilities: a CyberKnife® system, a dedicated medical linear accelerator (Novalis) equipped with m3 microMLC and circular cones, and an adaptive medical linear accelerator (Clinac 2100) equipped with an additional m3 microMLC. The second one was to determine the correction factors introduced in a recently proposed small field dosimetry formalism for different active detectors.Methods:Small field sizes were defined either by microMLC down to 6 × 6 mm2 or by circular cones down to 4 mm in diameter. OF measurements were performed with several commercially available active detectors dedicated to measurements in small fields (high resolution diodes: IBA SFD, Sun Nuclear EDGE, PTW 60016, PTW 60017; ionizing chambers: PTW 31014 PinPoint chamber, PTW 31018 microLion liquid chamber, and PTW 60003 natural diamond). Two types of passive dosimeters were used: LiF microcubes and EBT2 radiochromic films.Results:Significant differences between the results obtained by several dosimetric systems were observed, particularly for the smallest field size for which the difference in the measured OF reaches more than 20%. For passive dosimeters, an excellent agreement was observed (better than 2%) between EBT2 and LiF microcubes for all OF measurements. Moreover, it has been shown that these passive dosimeters do not require correction factors and can then be used as reference dosimeters. Correction factors for the active detectors have then been determined from the mean experimental OF measured by the passive dosimeters.Conclusions:Four sets of correction factors needed to apply the new small field dosimetry formalism are provided for several active detectors. A protocol for small photon beams OF determination based on passive dosimeters measurements has been recently proposed to French radiotherapy treatment centers.

Từ khóa


Tài liệu tham khảo

10.1088/0031‐9155/41/1/008

10.1088/0031‐9155/44/9/303

10.1118/1.1591431

10.1118/1.1536290

10.1118/1.2815356

10.1088/0031‐9155/47/11/401

10.1118/1.1567736

Derreumaux S., 2010, IAEA Proceedings of the International Symposium on Standards, Applications and QA in Medical Radiation Dosimetry, 273

2007, L'accident de surexposition au Centre Hospitalier Universitaire de Toulouse, Rapport d'expertise n°1, Vérification des protocoles expérimentaux d’étalonnage des microfaisceaux avant et après correction du dysfonctionnement

2008, Mesure de la dose absorbée dans les faisceaux de photons de très petites dimensions utilisés en radiothérapie stéréotaxique

IPEM, 2010, Small field MV photon dosimetry

10.1118/1.3005481

10.1118/1.2828195

10.1120/jacmp.v10i1.2939

10.1118/1.3660770

10.1088/0031‐9155/57/12/3741

10.1118/1.3658572

10.1016/j.radmeas.2009.12.005

10.1016/j.radmeas.2011.10.020

10.1118/1.1494830

10.3171/FOC‐07/10/EIntro

Technical Reference Guide Rev 1.0 BrainLAB Physics 2008.

International Organization for Standardization “Guide to the expression of uncertainty in measurement (GUM)” (1995).

10.1118/1.2734384

10.1118/1.2219774

10.1118/1.3404289

10.1016/j.radonc.2010.03.016

10.1118/1.2734383

10.1118/1.2975223

10.1088/0031‐9155/57/9/2587