Experimental assessment of the coincidence summing corrections in gamma-ray spectrometry of bulk samples

Applied Radiation and Isotopes - Tập 83 - Trang 1-7 - 2014
T. Boshkova1
1Faculty of Physics, St. Kliment Ohridsky, University of Sofia, 5 James Bourchier Blvd., 1164 Sofia, Bulgaria

Tài liệu tham khảo

Andreev, 1973, Determination of the γ-detection efficiency in energy peaks by means of nuclides having a complicated dacay scheme in close-geometry conditions, Izv. Acad. Nauk. SSSR, Ser. Fiz., 37, 1609 Blaauw, 2003, Cascade summing in gamma-ray spectrometry in marinelli-beaker geometries: the third efficiency curve, Nucl. Instr. Meth. A, 505, 311, 10.1016/S0168-9002(03)01075-1 Boshkova, 2003, Effective thickness of bulk samples in “close” measuring gamma-ray spectrometry, Appl. Radiat. Isot., 59, 1, 10.1016/S0969-8043(03)00125-8 Debertin, 1988 Debertin, 1979, Coincidence summing corrections in Ge(Li)-spectrometry at low source-to-detector distances, Nucl. Instr. Meth. A, 158, 471, 10.1016/S0029-554X(79)94845-6 Decombaz, 1992, Coincidence summing corrections for extended sources in gamma-ray spectrometry using Monte Carlo simulation, Nucl. Instr. Meth.A, 312, 152, 10.1016/0168-9002(92)90146-U De Felice, 2000, Fast procedures for coincidence-summing correction in γ-ray spectrometry, Appl. Radiat. Isot., 52, 745, 10.1016/S0969-8043(99)00239-0 Garcia-Torano, 2005, Monte Carlo calculations of coincidence-summing corrections for volume sources in gamma-ray spectrometry with Ge detectors, Nucl. Instr Meth. A, 544, 577, 10.1016/j.nima.2005.01.299 Gilmore, 1995 IAEA, 2002. Summary Report of the Proficiency Test for the Determination of Anthropogenic γ-Emitting Radionuclides in a Mineral Matrix, IAEA, Seibersdorf, 8 June, 2002. IEC 6 1452, 1995. Nuclear Instrumentation – Measurement of Gamma-ray Emission Rates of Radionuclides – Calibration and Use of Germanium Spectrometers. IEC 973 1989, Test Procedures for Germanium Gamma-Ray Detectors, June 1989. Lépy, 2010, Intercomparison of methods for coincidence summing corrections in gamma-ray spectrometry, Appl. Radiat. Isot., 68, 1407, 10.1016/j.apradiso.2010.01.012 Lépy, 2007, Total efficiency calibration for coincidence-summing corrections, Nucl. Instr. Meth. A, 579, 284, 10.1016/j.nima.2007.04.081 Lépy, 2006, Experimental validation of coincidence summing corrections computed by the ETNA software, Appl. Radiat. Isot., 64, 1340, 10.1016/j.apradiso.2006.02.042 Montgomery, 1995, A method for assessing and correcting coincidence summing effects for germanium detector efficiency calibrations, J. Radioanal. Nucl. Chem., 193, 71, 10.1007/BF02041918 Piton, 2000, Efficiency transfer and coincidence summing corrections for γ-ray spectrometry, Appl. Radiat. Isot., 52, 791, 10.1016/S0969-8043(99)00246-8 Sima, 2000, Accurate computation of coincidence summing corrections in low level gamma-ray spectrometry, Appl. Radiat. Isot., 53, 51, 10.1016/S0969-8043(00)00113-5 Vidmar, 2007, A method for calculation of true coincidence summing correction factors for extended sources, Appl. Radiat. Isot., 65, 243, 10.1016/j.apradiso.2006.07.012 Vidmar, 2006, Calculation of “LS-curves” for coincidence summing corrections in gamma ray spectrometry, Nucl. Instr. Meth.A, 556, 543, 10.1016/j.nima.2005.11.052